A simple laser ICP-MS ablation cell with wash-out time less than 100 ms
- In this paper a novel concept of ablation cell for laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) is presented. Suppression of the turbulence in the flush gas flow in the ablation region reduces the wash-out time of the ablation cell considerably. An ablation chamber which enables ICP-MS pulse duration down to several ten milliseconds has been designed. Dependence of the ICP-MS peak amplitude, width, and shape on the gas flow parameters is studied experimentally for aerosol ablated under laminar and turbulent conditions. Experiments demonstrate that the ICP-MS peak becomes sharper and the amplitude of the signal grows as the turbulence in the ablation cell is suppressed. Furthermore, the possibility of the LA-ICP-MS analysis with a sampling rate of more than 10 Hz has been demonstrated. Express in-depth profiling in the new ablation cell is demonstrated on examples of an Al -Zn multilayer structure and an industrial Mg -Zn coating.
Verfasserangaben: | E. L. Gurevich, R. Hergenröder |
---|---|
DOI: | https://doi.org/10.1039/b704700b |
ISSN: | 0267-9477 |
Titel des übergeordneten Werkes (Englisch): | J. An. At. Spectr. |
Publikationstyp: | Beitrag in einer (wissenschaftlichen) Zeitschrift |
Sprache: | Englisch |
Jahr der Fertigstellung: | 2007 |
Jahr der Erstveröffentlichung: | 2007 |
Datum der Freischaltung: | 08.11.2019 |
Jahrgang: | 22 |
Erste Seite: | 1043 |
Letzte Seite: | 1050 |
Fachbereiche: | Physikingenieurwesen (PHY) |
Publikationsliste: | Gurevich, Evgeny |
Lizenz (Deutsch): | ![]() |